Agilent 1100 Series G1365B Multi-Wavelength Detector You can add spectral acquisition to your HPLC analyses with the Agilent 1100 Series diode array detector and Verify separation quality with peak purity, multiple signals and spectral libraries – interactively or automatically in the same run as your analytical quantification
Agilent G1315B DAD Detector Manual
Type | Specification |
Detection type | 1024-element photodiode array |
Light Source | Deuterium and tungsten lamps |
Wavelength range | 190 – 950 nm |
Short term noise (ASTM) single and multi-Wavelength | ± 1 × 10-5 AU at 254 and 750 nm |
Drift | 2 × 10-3 AU/hr at 254 nm |
Linear absorbance range | > 2 AU (upper limit) |
Wavelength Accuracy | ± 1 nm |
Wavelength bunching | 1 – 400 nm |
Slit width | 1, 2, 4, 8, 16 nm |
Diode Width | < 1 nm |
Flow Cells | Standard: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi) pressure maximum Semi-Micro: 5 µl volume, 6 mm cell path length and 120 bar (1760 psi) pressure maximum High pressure: 1.7 µl volume, 6 mm cell path length and 400 bar (5880 psi) pressure maximum 500 nano: 0.5 µl volume, 10 mm cell path length and 50 bar (725 psi) pressure maximum |
Control and data evaluation | Agilent ChemStation for LC |
Analog outputs | Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs |
Communications | Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |
Safety and maintainence | Extensive diagnostics, error detection and display (through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas |
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter. |
Housing | All materials recyclable |